2013, 01, v.26 43-46
呼吸感应体积描记法校准方法在不同呼吸状态下的适用性
基金项目(Foundation): 首都医学发展科研基金(20073038)
邮箱(Email):
DOI: 10.16289/j.cnki.1002-0837.2013.01.012
摘要:

目的比较呼吸感应体积描记技术(RIP)的6种校准方法在人体不同呼吸状态下的适用性。方法用平静呼吸的6种校准参数测量平静和规范用力的每次呼吸通气量,用规范用力呼吸的6种校准参数测量规范用力的每次呼吸通气量,对比分析RIP系统与呼吸速度描记仪所测量的通气量值。结果 6种校准方法中,诊断性定性校准技术(QDC)适用于在平静呼吸下校准测量平静呼吸每次呼吸通气量值以及在规范用力呼吸下校准测量规范用力呼吸每次呼吸通气量值,LSQ改进算法(ILSQ)适用于在平静呼吸下校准测量规范用力呼吸每次呼吸通气量值。结论 RIP系统在平静呼吸状态下校准测量规范用力呼吸方式下每次呼吸通气量值方法可行。

Abstract:

Objective To compare the applicability of 6 calibration methods for respiratory volumes accuracy under different breathing conditions in respiratory inductive plethysmography (RIP).Methods Ventilation of each standardized forced breath (SFB) and natural breath(NB) employing parameters were measured with 6 methods under NB, and ventilation of each SFB employing parameters from 6 calibrated methods under SFB were evaluated.Ventilation measured from RIP system and pneumotachometer (PTM) were compared and analyzed.Results Qualitative diagnostic calibration(QDC) was appropriate for the two conditions of measuring ventilation of NB by calibrated parameter from the NB pattern and evaluating ventilation of SFB by calibrated parameter from the SFB pattern. Improved LSQ algorithm(ILSQ) was suitable for measuring ventilation of SFB by calibrated parameter from the NB pattern. Conclusion RIP system calibrated under NB is practical to measure ventilation of each breath under SFB.

参考文献

[1]朱元珏,陈文彬.呼吸病学[M].北京:人民卫生出版社,2002.

[2]朱蕾,刘又宁,于润江.临床肺功能[M].北京:人民卫生出版社,2004.

[3]郑劲平,陈蓉昌,钟南山.肺功能学:基础与临床[M].广州:广东科技出版社,2007.

[4]张政波,俞梦孙.背心式呼吸感应体积描记系统设计[J].航天医学与医学工程,2006,19(5):377-381.

[5]张劲农.非侵入性呼吸模式检测的基础和临床应用[J].国外医学呼吸系统分册,2005,25(8):630-633.

[6]张劲农,陶晓南等.呼吸电感应体积描记法用于睡眠气量测定的准确性研究[J].中华结合和呼吸杂志,1999,22(10):639-640.

[7]Konno K,Mead J.Measurement of the separate volume chan-ges of rib cage and abdom during breathing[J].Appl Physi-ol,1967,22:407-422.

[8]Sackner MA,Watson H,Belsito AS.et al.Calibration of re-spiratory inductive plethysmograph during natural breathing[J].Appl Physiol,1989,66:410-420.

[9]Poole KA,ThompsonJR,HallinanHM et al.Respiratory in-ductance plethysmography in healthy infants:a comparison ofthree calibration methods[J].Eur RespirJ,2000,16(6):1084-1090.

[10]Chadha TS,Watson H,Birch S,et al.Validation of respirato-ry inductive plethysmography using different calibration proce-dures[J].Am Rev Respir Dis,1982,125(6):644-649.

[11]Sartene R,Dartus C,BernardJL et al.Comparison of thora-coabdominal calibration methods in normal human subjects[J].J Appl Physiol,1993,75(5):2142-2150.

[12]Michael D,Sandra J.Comparison of calibration methods forrespiratory inductive plethysmography in infants[J].The A-merican Physiological Society,1987,63(5):1853-1861.

[13]Brown K,Aun C,Jackson E,et al.Validation of respiratoryinductive plethysmography using the ualitative Diagnostic Cali-bration me thod in anaesthetized infants[J].Eur Respir J1998,12:935-943.

[14]Anne de G,Manuel P,Yves V.Mathematical assessment ofqualitative diagnostic calibration for respiratory inductive ple-thysmography[J].J Appl Physio,2001,90:1025-1030.

[15]Poole KA,Thompson JR,Hallinan HM,et al.Respiratory in-ductance plethysmography in healthy infants:a comparison ofthree calibration methods[J].Eur Respir J,2000,16:1084-1090.

基本信息:

DOI:10.16289/j.cnki.1002-0837.2013.01.012

中图分类号:R56

引用信息:

[1]姬军,潘美玲,沙杭,等.呼吸感应体积描记法校准方法在不同呼吸状态下的适用性[J].航天医学与医学工程,2013,26(01):43-46.DOI:10.16289/j.cnki.1002-0837.2013.01.012.

基金信息:

首都医学发展科研基金(20073038)

检 索 高级检索

引用

GB/T 7714-2015 格式引文
MLA格式引文
APA格式引文